コンフォーカルワイヤーミオグラフ

コンフォーカルワイヤーミオグラフ
  • Product# DMT120CW  European Conformity

概要:

DMT120CWは レーザースキャニング顕微鏡で組織の蛍光イメージングを行いながらミオグラフ試験を実行できます。60μm以上の直径の血管/管状組織の研究に最適です。

コントロールユニットが付属し、アナログ出力が付いているのでPowerLabで血管の張力をモニタリングできます。

 

特徴:

  • シングルチャンネルワイヤーミオグラフ(10ml)
  • 倒立顕微鏡での開口部レンズのアクセスが容易なバスのデザインとなっています。
  • Reverse mounting によって、血管を垂直に設置することも可能。

仕様:

Technical Specifications

310A

410A

510A

610M

120CW

Tissue/Vessel size

>60 μm

>60 μm

>60 μm

>60 μm / >450 μm up to 10 mm

>60 μm

Chamber(s)

Single bath

Dual, independent or common

Dual, independent or common

Four individual chambers

Single bath/conical shape

Chamber material

Acid-resistant stainless steel

Acid-resistant stainless steel

Acid-resistant stainless steel

Acid-resistant stainless steel

Acid-resistant stainless steel

Chamber volume

Max. 10 ml

Max. 10 ml

Max. 10 ml

Max. 8 ml

Max. 10 ml

Chamber suction

No

No

No

Manual or automatic, time controlled, user defined

No

Chamber cover

Yes

With connection for suction and gassing

With connection for suction and gassing

Supplied with connections for gassing

With connection for suction and gassing

Force range

-100 -/+200 mN

-100 - /+200 mN

-100 - /+200 mN

User selectable at +/-200/400/800/1600 mN

-100 - /+200 mN

Force resolution

0.01mN

0.01mN

0.01mN

0.01mN

0.01mN

Micropositioners

Manually operated

Manually operated

Automatical linear motor operated

Manually operated

Manually operated precision micrometer

Weight calibration

Manual

Manual

Manual

Semi Automatic

Manual

Heating

Built into chamber,
independent of superfusion
Built into chamber,
independent of superfusion
Built into chamber,
independent of superfusion
Built into chamber,
independent of superfusion
Built into chamber,
independent of superfusion

Temp. range

Ambient temp. 
to 50 °C
Ambient temp. 
to 50 °C
Ambient temp. 
to 50 °C
Ambient temp. 
to 50 °C
Ambient temp. 
to 50 °C

Temp. resolution

0.1 °C

0.1 °C

0.1 °C

0.1 °C

0.1 °C

Temp. probe

External

External

External

External

External

Output reading

Force (mN)

Force (mN)

Force (mN)

Force (mN)

Force (mN)

Analog output

Up to four output, 1.0V full scale for all acquired signals, user defined
Up to four output, 1.0V full scale for all acquired signals, user defined
Up to four output, 1.0V full scale for all acquired signals, user defined

Independent filtered 4 channel output at 1.0V full scale

Up to four output, 1.0V full scale for all acquired signals, user defined

Serial output

Serial interface - RS232/RS485

Serial interface - RS232/RS485

Serial interface - RS232/RS485

Serial interface - RS232/RS485

Serial interface - RS232/RS485

Voltage

100 to 240 VAC (auto) 50/60Hz

100 to 240 VAC (auto) 50/60Hz

100 to 240 VAC (auto) 50/60Hz

100 to 240 VAC (auto) 50/60Hz

100 to 240 VAC
(auto) 50/60Hz

Ambient temp.

15-30 °C

15-30 °C

15-30 °C

15-30 °C

15-30 °C

Optional accessories






- range

pH 0 - 14

pH 0 - 14

pH 0 - 14


pH 0 - 14

- temp. correction

0 - 50 °C

0 - 50 °C

0 - 50 °C


0 - 50 °C